				The CS215-PWS is a temperature and relative humidity probe with a special connector on the cable for attaching to a PWS100. It uses a Swiss-made digital humidity and temperature element that makes accurate and stable measurements. The CS215-PWS outputs an SDI-12 signal measurable by most Campbell Scientific dataloggers.
Read MoreFor this version of the CS215 Temperature and RH Probe, the cable has, instead of pigtails, a connector for attachment to the PWS100. Cable length is 1 meter.
The CS215 uses the Sensirion SHT75, a combined relative humidity and temperature element, to provide accurate, stable measurements. The Sensirion SHT75 element is field-replaceable, eliminating the downtime typically required for the recalibration process. The CS215 outputs an SDI-12 signal that's measurable by many Campbell Scientific dataloggers.
The CS215 should be housed in a solar radiation shield—typically the 41303-5A. The 41303-5A 6-plate naturally aspirated shield attaches to a mast, crossarm, or tower leg.
 
				For this version of the CS215 Temperature and RH Probe, the cable has, instead of pigtails, a connector for attachment to the PWS100. Cable length is 1 meter.
| Sensing Element | Sensirion SHT75 | 
| Communication Standard | SDI-12 V1.3 (responds to a subset of commands) | 
| Supply Voltage | 6 to 16 Vdc (typically powered by data logger’s 12 V supply) | 
| Typical Current Drain | 
 | 
| EMC Compliance | Tested and conforms to IEC61326:2002. | 
| Operating Temperature Range | -40° to +70°C | 
| Diameter | 
 | 
| Probe Length | 18.0 cm (7.1 in.) including strain relief | 
| Cable Length | 10 m (33 ft) | 
| Weight | 350 g (12.3 oz) with 10-m (33-ft) cable | 
| Relative Humidity | |
| Measurement Range | 0 to 100% RH (-20° to +60°C) | 
| Output Resolution | 0.03% RH | 
| Accuracy | 
 | 
| Short-Term Hysteresis | < 1% RH | 
| Temperature Dependence | Better than ±2% (-20° to +60°C) | 
| Typical Stability | ±1.0% per year | 
| Response Time with Filter | < 20 s (63% response time in still air) | 
| Calibration Traceability | NIST and NPL standards | 
| Temperature | |
| Measurement Range | -40° to +70°C | 
| Output Resolution | 0.01°C | 
| Accuracy | 
 | 
| Response Time with Filter | < 120 s (63% response time in air moving at 1 m s-1) | 