The Half Bridge instruction should be selected (not the 3 Wire or 4 Wire options) in Short Cut. The following parameters are data logger specific, but the following can be used on the CR1000 datalogger with the sensor wired to SE1, and Vx1:
| How many HalfBr Sensors? | 1 |
| Total Bridge Resistance | 1000 ohm |
| Excitation Voltage | 2500 mV |
| Sensors per excitation channel | 1 |
| Range of sensor voltage | 2500 mV |
| Reverse Excitation | Yes |
| Measurement Integration |
50 or 60Hz |
| Settling Time | 20000 µs |
| Multiplier | Determined during installation (see manual) |
| Offset | Determined during installation (see manual) |